| model | EHR Standard new | EHR High Precision new |
|---|---|---|
| ch. | 4 | 4 |
| Vnom | 2kV - 6kV | 2kV - 6kV |
| Vin | 24V | 24V |
| ripple & noise | 10mV | 2mV - 3mV |
| HV connector | SHV | SHV |
| control | iCS | iCS |
| options | INH L TC SLA SLP | INH L TC SLA SLP |
| manual | EHR User Guide | EHR User Guide |
| datasheet | see configurations | see configurations |
| image | ![]() | ![]() |
EHR
Polarity switchable High Precision HV Module
The new EHR series represents a system capable multichannel high voltage module - equipped with the finest iseg HV generation technology in MMS system (Eurocard format). The module is made of high-precision components such as 24 bit ADC and up to 20 bit DAC and provides comprehensive safety features.
The EHR provides 4 HV-channels, each with an independent voltage and current control and reversible polarity.
A flexible 6kV channel provides a maximum versatility: With three electronically switchable HV-generation modes it can supply 4 mA up to voltages of 2 kV, 3 mA up to 4 kV or 2 mA up to 6 kV. Alternatively the EHR can be equipped with cost efficient 2kV/6mA fixed channels.
By offering different configurations and options this module perfectly covers various types of applications such as detector supply, experimental setup or lab use.
- 4 channels
- Electronically switchable polarity for each channel independently
- 2 kV / 6 kV versions
- 6 kV channel with electronical switchable modes: up to 2 kV/4 mA, 4 kV/3 mA or 6 kV/2 mA
- High precision / very low ripple and noise
- Second current measurement range 20 µA for high precision version
- Voltage and current control per channel
- Hardware voltage and current limits
group | model | Vnom | Inom | Pnom/ch | ch. | polarity | ripple & noise | interfaces | HV connector | options | image | manual |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| EHR Standard new | EHR 40 20 itemcode: ER040020r605 configure now | 2kV | 6mA | 12W | 4 | electronically switc... | 10mV | CAN | SHV | INH L TC SLA SLP | ![]() | EHR User Guide |
| EHR Standard new | EHR 40 60 itemcode: ER040060r405 configure now | 6kV | 4mA | 12W | 4 | electronically switc... | 10mV | CAN | SHV | INH L TC SLA SLP | ![]() | EHR User Guide |
| EHR High Precision new | EHR 42 20 itemcode: ER042020r605 configure now | 2kV | 6mA | 12W | 4 | electronically switc... | 2mV | CAN | SHV | INH L TC SLA SLP | ![]() | EHR User Guide |
| EHR High Precision new | EHR 42 20 itemcode: ER042020r104 configure now | 2kV | 100µA | 0.2W | 4 | electronically switc... | 2mV | CAN | SHV | INH L TC SLA SLP | ![]() | EHR User Guide |
| EHR High Precision new | EHR 42 60 itemcode: ER042060r405 configure now | 6kV | 4mA | 12W | 4 | electronically switc... | 3mV | CAN | SHV | INH L TC SLA SLP | ![]() | EHR User Guide |
| EHR High Precision new | EHR 42 60 itemcode: ER042060r104 configure now | 6kV | 100µA | 0.6W | 4 | electronically switc... | 3mV | CAN | SHV | INH L TC SLA SLP | ![]() | EHR User Guide |
| code | description |
|---|---|
| INH | INHIBIT |
| L | low current output |
| TC | improved temperature coefficient |
| SLA | active safety loop |
| SLP | internally powered safety loop |